2.3  CALIBRATION PROCEDURES 
2.3.1  CALIBRATION EQUIPMENT 
Calibration requires that a calibration system is used, i.e., equipment that applies accurate voltage, load current 
and load angle to the unit being calibrated, while measuring the response from the unit being calibrated in a re-
peatable way. By repeatable we mean that the calibration system is synchronized to the meter being calibrated. 
Best results are achieved when the first pulse from the meter opens the measurement window of the calibration 
system. This mode of operation is opposed to a calibrator that opens the measurement window at random time 
and that therefore may or may not catch certain pulses emitted by the meter. 
It is essential for a valid meter calibration to have the voltage stabilized a few seconds be-
fore the current is applied. This enables the Demo Code to initialize the 71M6541F and to 
stabilize the PLLs and filters in the CE. This method of operation is consistent with meter 
applications in the field as well as with metering standards.  
During calibration of any phase, a stable mains voltage has to be present on phase A. This 
enables the CE processing mechanism of the 71M6541F necessary to obtain a stable cali-
bration. 
2.3.2  PHASE-BY-PHASE CALIBRATION 
Each meter phase must be calibrated individually. Some calibration systems do not allow selective control of 
currents in each phase. Each phase can still be individually calibrated using the following sequence: 
•  When calibrating phase A, the calibration coefficient for the current in phase B is set to zero. This way, 
the pulses are generated solely based on phase A. The kH factor of the calibration system must be ad-
justed by -50% to account for the suppression of 50% of the energy. 
•  When calibrating phase B, the calibration coefficient for the current in phase A is set to zero. This way, 
the pulses are generated solely based on phase B. The kH factor of the calibration system must be ad-
justed by -50% to account for the suppression of 50% of the energy. 
•  For the final step, both current calibration coefficients are set to their calibration values and the meter 
can be tested at the original kH setting. 
2.3.3  DETAILED CALIBRATION PROCEDURES 
The procedures below show how to calibrate a meter phase with either three or five measurements. The 
PHADJ equations apply only when a current transformer is used for the phase in question. Note that positive 
load angles correspond to lagging current (see Figure 2-6). 
Figure 2-6: Phase Angle Definitions 
Voltage
Current
+60°
Using EnergyGenerating Energy
Current lags
voltage
(inductive
)
Current leads
voltage
(capacitive
)
-60°
Voltage
Positive
direction