Instruction Manual

6612_OMU_S2+2_URT_V1_14 Firmware Description Document UG_6612_017
14 Rev. 1.0
4.5.2 Atomic Calibration Commands
The atomic calibration commands provide individual calibration of voltage, current, temperature, watts
and a sequence of these results in providing full calibration for the unit.
4.5.2.1 CLV Command
An example of an atomic calibration command would be to calibrate voltage with the CLV command. The
CLV command calibrates voltage to the target value and tolerance and saves the coefficients to flash.
The CLV command example is given below:
>CLV<CR>
The response is:
VCal OK
>
4.5.2.2 CLI Command
The user can then calibrate the current on Outlet 1 using the CLI1 command. The CLI1 command
calibrates the current on Outlet 1 to the target value and tolerance and saves the coefficients to flash.
The CLI1 command example is given below:
>CLI1<CR>
The response is:
ICal 0 OK
>
4.5.2.3 CLP Command
The user can calibrate for phase added by a current transformer by using the CLP command. The CLP
command calibrates the phase on Outlet 1 to the target value and tolerance and saves the coefficient to
flash. An example of the procedure is given below.
Apply a controlled precision voltage and current signal at a set phase angle.
1. Enter target phase angle at )C3.
2. Enter phase tolerance at )BF
3. Enter phase calibration command.
>CLP<CR>
The response is
>PCal 1 OK