• EEPROM HW error (E101)
The relay remains picked up even after acknowledging.
• Implausible calibration data (E103)
The relay remains picked up even after acknowledging.
• Bus error reading the min/max data after power-up (E104)
The relay remains picked up even after acknowledging.
• Bus error reading the relay data after power-up (E105)
The relay remains picked up even after acknowledging.
• Universal card HW error (E106)
The relay remains picked up even after acknowledging.
• Pulse buffer overflow (E210)
The relay drops out after acknowledging.
• Pump error at the digital input x in question (E22x)
The relay remains picked up even after acknowledging.
Delay
G09-RIA452ZZ-15-00-xx-en-003
Fig. 18: Delay
The following parameters must be configured:
Menu Function (menu item) Setting value
LIMIT 1 to 8/M10 to 17 Setpoint A
Hysteresis
Delay
Value for threshold
Value for hysteresis
Delay time in [s]