User Manual

MCR METER Manual
7
Vm ON/OFF :Open or close voltage monitoring function
Im ON/OFF :Open or close current monitoring function
3.2 < BIN DISP > (MCR5010 no this feature)
This page can be set as follows:
COMP ON/OFF :to open or close the comparison function
AUX ON/OFF :to open or close AUX.
AUX is the main parameter is qualified, and the Deputy parameter is not qualified.
COUT ON/OFF :to open or close the counting function
NOMINAL By using digital key and softkey input. (the reference value for the
comparison of the main parameters)
MODE Δ% / ABS : to choice the comparison mode.
Δ% :Relative percentage mode
ABS :Absolute value comparison mode
LOW and HIGH :Through digital key, the OK key and the softkey input.
Test result display area:DISP ON/OFF;DEV OFF/ABS/%(Same as in 3.1)
3.3 <MULTI PARA.DISP> (only MCR5200 has this function)
This page can be set as follows:
FREQ : Set frequency
LEVEL: Set test voltage
1/2/3/4/5 test parameter: The test parameters required by soft region selection.
3.4 <MEAS SETUP>
This page can be set as follows:
ResSource :30Ω/100Ω,Select output resistance mode
AVERAGE :1~255.By using digital key input or softkey "INCR+" and "DECR -".
RangeDwell Range delay. 0~6000ms.By using digital key input or softkey "INCR+" and "DECR
-".
TrigSource :INT,EXT.(Trigger mode)
TrigDelay :0~6000ms.By using digital key input or softkey "INCR+"
and "DECR -".
TrigEdge :RISING,FALLING .The rising edge of the trigger; the falling
edge of the trigger
Handler :CLEAR,HOLD,PULSE. Sorting signal mode
PulseWidth :1~9999ms.(pulse width)
3.5 <SYSTEM SETUP>
This page can be set as follows:
ThemeTRAD-BLUE,TRAD BLACK,BLACKISH GREEN.(Display style)
Language :English、中文 . (Language environment)
Key Tone :ON,OFF. Open or close button sound
ToneSour :MASTER,EARPHONE,ALL.(Buzzer mode,EARPHONE socket is optional)
PassTone :OFF,LONG,SHORT,TWO SHORT
FailTone :OFF,LONG,SHORT,TWO SHORT
ParaSave :AUTO SAVE,AUTO LOAD,NO SAVE
AUTO SAVE : This feature will save the settings of the measurement parameters in real time,
the next boot state will be the same as the last shutdown.
AUTO LOAD :Automatically invokes the last loaded internal file.