User Manual
MCR METER Manual
17
Select "YES", the instrument will enter a short circuit cleared process, if the short circuit failure,
will jump to the second step to re ask;Note: When performing SHORT operation, must
ensure that the test fixture is short.
4. “To make UNKNOWN open”is displayed on the screen.At this time, there are 3 regional soft
options: [YES] ,[NO], [CANCEL].
5. Select [CANCEL], then quit the operation of OPEN;
Select "NO", skip the OPEN operation, and enter the sixth step;
Select "YES", the instrument will enter a open circuit cleared process, if the open circuit failure,
will jump to the second step to re ask;Note: When performing OPEN operation, must
ensure that the test fixture is open.
6. Clear end.
Another way to clear:(In this way, the instrument will clear current frequency point)
Figure 6-2
1. Press [TOOL] key in <MEAS DISP> page or in <BIN DISP> page ,or in <MULTI PARA. DISP>
page, the [SINGLE SHORT] and [SINGLE OPEN] will displayed .
2. If you need SHORT operation, please make the test fixture short, and then select the [SINGLE
SHORT] softkey, the current test status will be short circuit cleared.
3. If you need OPEN operation, please make the test fixture open, and then select the [SINGLE
OPEN] softkey, the current test status will be open circuit cleared.
6.2 Set sorting
Figure 6-2 <BIN DISP> page
1. Press the "MEAS" button on the panel to enter the < MEAS DISP > page (as shown in Figure
6-1), and press "BIN DISP " softkey to enter < BIN DISP > page (as shown in Figure 6-2).
2. Move the cursor to the "NOMINAL", press the number keys and softkey to input the nominal
value of main parameter.
3. Move the cursor to the lower or upper limit of the BIN1~ BIN3, use digital keys and OK key
to enter the upper and lower limits.
4. Move the cursor to the lower or upper limit of 2nd ,press the number keys and softkey to
input the upper and lower limits of the second parameter.