MCR METER Manual MCR5000 SERIES USER MANUAL MCR5010 MCR5030 MCR5100 MCR5200 MATRIX TECHNOLOGY INC.
MCR METER Manual Chapter 1 Instrument inspection and preparation ------------------------------------2 1.1 The crates ------------------------------------------------------------------2 1.2 Power connection ------------------------------------------------------------3 1.3 Fuse and toggle switch --------------------------------------------------------3 1.4 Environmental requirements ---------------------------------------------------3 1.
MCR METER Manual Chapter 1 Instrument inspection and preparation Please be sure to check the receipt of the instrument, and must understand and have the condition before using the instrument. 1.1 The crates After receiving the product, it is found that the packing is badly damaged. Please keep it until it is normal for the whole machine and the accessories. Check the machine, if the instrument is damaged, please contact the dealer or the company in charge of the business.
MCR METER Manual is in good contact. The test fixture or test cable is connected to the HD, HS, LS and LD of the front panel of the instrument. The shield layer can be connected to the rear panel of the instrument, and the shield layer can be connected with the rear panel of the instrument. User homemade or other company test fixtures or test cables may result in incorrect measurement results. 1.
MCR METER Manual 2.1 Front panel description 6 7 9 8 10 PASS FAIL 11 12 13 14 15 MEAS SETUP SYSTEM FILE 0 1 4 7 +/- 2 3 ABC DEF 5 6 GHI JKL MNO PQRS 8 1. 3. 5. 7. 8. 9. 10. 11. 12. 13. 14. 15. 17. 18. 19. 2 3 4 WXYZ 5 Power switch 2. Instrument type label USB interface 4. Tool key Test terminal:LD、LS、HS、HD 6. 4.
MCR METER Manual 2.3 Basic operation 1. 2. 3. 4. Direction key: used to move the cursor. Soft keys, used to change the parameters set of features. Shortcut menu: MEAS, SETUP, SYSTEM, FILE, press directly into the < MEAS DISP >, < MEAS SETUP >, < SYSTEM SETUP >, < INTER/EXT Files List> function page settings. Reminder: when in < MEAS DISP > page, and the cursor in < MEAS DISP >, press MEAS key, < MEAS DISP > page test results will be full screen display.
MCR METER Manual Rs: Series equivalent resistance(ESR)) Rp: Parallel equivalent resistance(EPR) D : loss Q : quality factor R : resistance X : reactance Zs: Series equivalent impedance θ: angle r : angle RANGE:AUTO,HOLD,INCR +,DECR FREQ:Set frequency MCR5010 : 100Hz,120Hz,1kHz,10kHz, MCR5030:100Hz,120Hz,1kHz,10kHz,20kHz,30kHz MCR5100: 50Hz,60Hz,100Hz,120Hz,1kHz,10kHz, 20kHz,40kHz,50kHz,100kHz MCR5200: 40Hz、50Hz、60Hz、80Hz、100Hz、120Hz、150Hz、200Hz、250Hz、300Hz、 400Hz、500Hz、600Hz、800Hz、1kHz、1.2kHz、1.5kHz、2kHz、2.
MCR METER Manual Vm ON/OFF Im ON/OFF :Open or close voltage monitoring function :Open or close current monitoring function 3.2 < BIN DISP > (MCR5010 no this feature) This page can be set as follows: COMP ON/OFF :to open or close the comparison function AUX ON/OFF :to open or close AUX. AUX is the main parameter is qualified, and the Deputy parameter is not qualified. COUT ON/OFF :to open or close the counting function NOMINAL :By using digital key and softkey input.
MCR METER Manual NO SAVE : Each boot display parameters for the default configuration of the instrument. PassWord :OFF,LOCK SYSTEM,LOCK FILE,MODIFY,SAVE TO U_DISK OFF : No password LOCK SYSTEM : File and boot need password to perform the operation LOCK FILE : File requires a password to perform the operation MODIFY : Modify password. Factory default password : 0010(MCR5010) 0100(MCR5100) 0200(MCR5200) SAVE TO U_DISK : Save the password to U disk.For example, the password file named "0010.
MCR METER Manual The parameters set by the user in form of a file stored in the instrument internal non-volatile memory, when the next time you want to use the same settings, users do not need to set these parameters, only need to load the appropriate file, the last set of parameters are obtained. Thus greatly saving the time for the user to re set the parameters and improving the production efficiency. Press FILE button to enter the < INTER Files List > page.
MCR METER Manual 4.1 Range In 100 ohm source resistance, using a total of 5 range: 31.6Ω,100Ω,1kΩ,10kΩ,100kΩ In 30 ohm source resistance, using a total of 6 range : 10Ω,30Ω,100Ω,1 kΩ,10 kΩ,100 k Ω The effective measurement range of each range is shown in the following table. Table 4-1 The range in 100Ω source resistance No. Range Effective Resistance measurement range 0 100 kΩ 100kΩ-100MΩ 1 10 kΩ 10kΩ-100kΩ 2 1 kΩ 1kΩ-10kΩ 3 100Ω 50Ω-1kΩ 4 30Ω 0Ω-50Ω Table 4-2 The range in 30Ω source resistance No.
MCR METER Manual Accuracy of D The accuracy of De about D is expressed by the following formula: De= Ae 100 When Dx≤0.1, De = De When Dx>0.1 ,De = De *(1+Dx) Accuracy of Q The accuracy of Qe about Q is expressed by the following formula: Qe = Qx De 1 Qx De Here, Qx is the test value of Q. De is the accuracy of D It is effective,when Qx×De<1 Accuracy ofθ The accuracy of θ is expressed by the following formula: θe = 180 Ae 100 [deg] Accuracy of RP When Dx≤0.
MCR METER Manual 10n [S] 100M 100kH 10pF [OHM] 1pF 10kH 100fF 1kH 10fF 100H 100pF 100n 10H 10M 1nF 1.5M 1u 1H 1M 10nF 10u 100mH 100k 100nF 1m 10mH 10k |Z|,R,X |Y|,G,B 100u 1uF 1mH 1k 0.1 0.2 10uF 10m 100uH 100 100uF 100m 10uH 10 1mF 1.5 1 1uH 1 0.25 0.35 10mF 10 100nH 100m 0.65 1.0 100mF 100 10m 10nH 50Hz 100Hz 1kHz 10kHz 100kHz 200kHz Figure 4-1 Basic measurement accuracy: A Figure 4-1, in the junction line, can choose a smaller value.
MCR METER Manual spee d MED, SLO W frequency Ka(Zm < 500) Kb(Zm > 500 ) fm100Hz ( 1 10 3 200 100 )(1 )(1 ) Zm Vs fm Z m (110 9 )(1 100Hzfm ≤100kHz fm100kH z fm100Hz ( 70 100 )(1 ) Vs fm 110 3 200 ( )(1 ) Zm Vs Z m (1 10 9 )(1 70 ) Vs 1 10 3 200 ( )(2 ) Zm Vs Z m (3 10 9 )(1 70 ) Vs 2.5 10 3 400 100 )(1 )(1 ) Zm Vs fm Z m (2 10 9 )(1 100 100 )(1 ) Vs fm 100Hzfm 2.
MCR METER Manual realize the automatic sorting test. Table 5-1 Pin definition table for Handler interface Pin Signal Described number name 1 /P1 Sorting result output. 2 /P2 All signal with pull up resistor collector output. Up pull supply can 3 /P3 choose the internal +5V power supply, or external power supply 10 /NG EXTV (factory default). 11 /AUX The internal pull resistor is 4.7kΩ. 12,13 /EXT.TRIG The rising edge of this signal triggers the instrument to measure.
MCR METER Manual Figure 5-1 Handler interface pin definition Figure 5-2 Timing diagram of interface signal Description: Time Minimum value 1us 200us 0us T1 : Trigger pulse width T2 : Measurement start delay time T3 :Trigger wait time after /EOT output 15 Maximum value Display time + 200us ---
MCR METER Manual Figure 5-3 Schematic diagram of Handler interface signal When the internal power supply is used, the 1 and 2 of the J205 are connected; the 1 and the 2 of the J204 are connected. When using an external power supply (factory default), the 2 and 3 of the J205 are connected; the 2 and 3 of the J204 are connected. Chapter 6 Operation example 6.1 Clear operation One way to clear:(In this way, the instrument will clear all the frequency points) Figure 6-1 Measurement display page Step: 1.
MCR METER Manual Select "YES", the instrument will enter a short circuit cleared process, if the short circuit failure, will jump to the second step to re ask;Note: When performing SHORT operation, must ensure that the test fixture is short. 4. “To make UNKNOWN open”is displayed on the screen.At this time, there are 3 regional soft options: [YES] ,[NO], [CANCEL]. 5.
MCR METER Manual 5. Move the cursor to "COMP", open the comparison switch. Appendix: list of parameters of various models Measurement function MCR test parameters | Z|, |Y|, C, L, X, B, R, G, D, Q, θ MCR5010 :100Hz,120Hz,1kHz,10kHz MCR5030 :100Hz,120Hz,1kHz,10kHz,20kHz,30kHz Test frequency MCR5100 :40Hz、50Hz、60Hz、80Hz、100Hz、120Hz、150Hz、200Hz、 250Hz、300Hz、400Hz、500Hz、600Hz、800Hz、1kHz、 1.2kHz、1.5kHz、2kHz、2.5kHz、3kHz、4kHz、5kHz、6kHz、 7.
MCR METER Manual Comparator function storage Interface Five BIN sorting function(BIN1,BIN2,BIN3,AUX,OUT) Internal: 105; external can save 500 Standard:RS232C,USB HOST,HANDLER(Except MCR5010 )、 Optional:USB DEVICE,headphone jack,Foot switch interface ,GPIB 19