Operator's Manual

10 | P a g e
© ExSens TECHNOLOGY (PTY) LTD2017
BATT VOLTAGE: PASS / FAILED
Once the self-test has completed the instrument will display the company logo as the start-up screen.
Figure 3: LOGO Display screen
Once the initialization is complete, the instrument will switch to the operational screen. While in the
warm-up period the readings will be blanked off. Once warm-up has completed the readings will be
active.
Figure 4: Normal Display
The layout of the display is as follows:
The model of Instrument is displayed on top of screen as shown in Figure 4, this will always be ExSens
X4. (Model of this instrument).
The instrument is capable of detecting up to four different gasses. The Flammable gas will always be a
pellistor and/or InfraRed sensor capable of detecting flammable gas in the range of 0.00% to 5.00%
vol/vol. The display can also be configured to display the value from 0.0% to 99.0%LEL.
The instrument also accommodates two toxic sensors that have various ranges depending on the type
of sensor and exposure limits harmful to humans. The measurement ranges are from 0ppm to
999ppm.