User guide

76
GlobalFiler
Express PCR Amplification Kit User Guide
Appendix D PCR Work Areas
Amplified DNA work area
D
Tube decapper, autoclavable
•Vortex
Amplified DNA work area
IMPORTANT! Place the thermal cyclers in the Amplified DNA Work Area.
You can use the following systems:
•Veriti
®
96-Well Thermal Cycler (Part no. 4375786)
GeneAmp
®
PCR System 9700 with the Silver 96-Well Block
GeneAmp
®
PCR System 9700 with the Gold-plated Silver 96-Well Block
IMPORTANT! The GlobalFiler
Express Kit is not validated for use with the
GeneAmp
®
PCR System 9700 with the Aluminium 96-Well Block. Use of this
thermal cycling platform may adversely affect performance of the GlobalFiler
Express Kit.
IMPORTANT! The GlobalFiler
Express Kit is validated for use with the standard
Veriti
®
96-well Thermal Cycler (Part no. 4375786) NOT the Veriti
®
96-Well Fast
Thermal Cycler (Part no. 4375305). Please ensure you are using the correct Veriti
®
Thermal Cycler model.