User guide

56
GlobalFiler
Express PCR Amplification Kit User Guide
Chapter4 Analyze Data
Set up GeneMapper
®
ID-X Software for data analysis
4
SQ & GQ tab
settings
IMPORTANT! The values shown are the software defaults and are the values we used
during developmental validation. Perform appropriate internal validation studies to
determine the appropriate values to use.
Set the ACC GQ Weighting according to your laboratory's use of the ACC PQV. For
example, set the ACC GQ Weighting to 0.3 or higher to flag samples in which the
Amelogenin result is anything other than X, X or X, Y, or does not agree with the
results for the DYS391 or the Y indel markers.
Create a size
standard
The size standard for the GeneScan
600 LIZ
®
Size Standard v2.0 contains the
following peaks: