User guide

50
GlobalFiler
Express PCR Amplification Kit User Guide
Chapter4 Analyze Data
Set up GeneMapper
®
ID-X Software for data analysis
4
d. Select AmpFLSTR_Stutter_v3X.txt, then click Import.
Note: Importing this file associates the marker stutter ratio with the bin set
in the AmpFLSTR_Panels_v3X folder and overwrites any existing stutter
ratios associated with the panels and bins in that folder.
9. View the imported marker stutters in the navigation pane:
a. Double-click the AmpFLSTR_Panels_v3X folder to display its list of kits in
the right pane.
b. Double-click the GlobalFiler_Express_v1.1X folder to display its list of
markers below it.
c. Double-click D21S11, then click Stutter Ratio & Distance to display the
Stutter Ratio & Distance view for the marker in the right pane.
10. Click Apply, then OK to add the GlobalFiler
Express Kit panel, bin set, and
marker stutter to the GeneMapper
®
ID-X Software database.
IMPORTANT! If you close the Panel Manager without clicking Apply, the panels,
bin sets, and marker stutter will not be imported into the GeneMapper
®
ID-X
Software database.