User guide

18
GlobalFiler
Express PCR Amplification Kit User Guide
Chapter1 Overview
Materials and equipment
1
Figure2 Spectral calibration of the six dyes used in the GlobalFiler
Express Kit
Materials and equipment
Kit contents and
storage
The GlobalFiler
Express Kit contains sufficient quantities of the following reagents to
perform 200 (Part no. 4476609) or 1000 (Part no. 4474665) amplifications at
15 µL/amplification:
IMPORTANT! The fluorescent dyes attached to the primers are light-sensitive. Protect
the primer set, amplified DNA, allelic ladder, and size standard from light when not in
use. Keep freeze/thaw cycles to a minimum.
Table2 Kit Contents and Storage
LIZ
®
TAZ
VIC
®
6-FAM
SID
NED
Component Description 200 reaction 1000 reaction Storage
GlobalFiler
Express Master Mix
Contains enzyme, salts, dNTPs,
bovine serum albumin, enzyme,
and 0.05% sodium azide in
buffer and salt.
1 tube, 1.2 mL 1 bottle, 6 mL –15 to –25°C upon
receipt, 2 to 8°C after
initial use for up to
1month
GlobalFiler
Express Primer Set
Contains forward and reverse
primers to amplify human DNA
targets.
1 tube, 1.2 mL 1 bottle, 6 mL
GlobalFiler
Express Allelic
Ladder
Contains amplified alleles.
See Table1 on page 13 for a list
of alleles included in the allelic
ladder.
1 tube, 0.065 mL 1 tube,
0.15 mL
DNA Control 007 Contains 2 ng/µL human male
genomic DNA in 0.05% sodium
azide and buffer
.
See Table1 on page 13 for
profile.
1 tube, 0.05 mL 2 tubes,
0.05 mL/tube
The DNA Control 007 is included at a concentration appropriate to its intended use as an amplification control (i.e., to provide confirmation of
the capability of the kit reagents to generate a profile of expected genotype). The DNA Control 007 is not designed to be used as a DNA
quantitation control and laboratories may expect to see variation from the labelled concentration when quantitating aliquots of the DNA Control
007.