LEDs indication wireless spirometry sensor unit:
LED Indication Description Action
Orange LED on Charging in cradle
Orange LED off No charging/charging complete
in cradle
Orange LED blinking
slowly
Low battery Put the wireless sensor unit into charging
station.
Orange LED blinking
fast
Indicates an error of wireless
spirometry sensor unit
Indicates an error in the Wireless Spirometry
Sensor Unit.
- Restart the wireless sensor unit by pressing
the power button on the wireless sensor
unit for > 8 seconds
- If issue persists after restart, check in
MasterScope “Device Management” if
sensor is connected properly and
reconnect if necessary.
Blue LED on Wireless Spirometry Sensor
Unit is transferring data to
MasterScope
Blue LED off Wireless Spirometry Sensor Unit
in sleep mode or turned off
Press the power on button on the Wireless
Spirometry Sensor Unit to turn it on or
in case the battery is discharged, put the
sensor in the Charging Station
Blue LED blinking
slowly
Wireless Spirometry Sensor Unit
turned on and connected to
MasterScope
LEDs indication wireless spirometry sensor unit:
LED Indication Description Action
Blue LED on Charging station ready
Blue LED off No power Ensure that Power supply is plugged in and
USB connector properly attached on the
backside of the charging station.
The orange LED has priority before the blue LED!