Use Instructions
Page 88/233
MasterScope - Instructions for Use
Version 03.00 • Date 19FEB2020
Click <OK> and restart the linearity check with <F1>. If the check continues to fail,
try changing the pneumotach head, and repeat the linearity check. If this also fails,
please contact ERT Customer Care.
The linearity check log report can be printed by selecting <Calibration> and the
<Linearity Check Log> icon.
If any of the strokes are outside this limit, the calibration check will automatically
stop, and the message “Measured volume is outside
+/- 3.5% of the pump volume.“