Use Instructions
Page 84/233
MasterScope - Instructions for Use
Version 03.00 • Date 19FEB2020
3. Press <F1> to start the procedure. An automatic zero adjustment of the
connected pneumotach will be performed. The following window appears:
4. Press <OK> to start zero adjustment.
5. The screen now shows you the standard flow/volume calibration graph.
6. Start to pump the calibration syringe with complete strokes - incomplete
strokes could lead to errors.
7. The first three strokes are discarded and can be performed at any flow. A
counter in the bottom right window will count down the number of strokes
performed. The inhaled (IN) and exhaled (EX) volume of each stroke is
recorded in the top right window.