Click <OK> and restart the calibration check with <F1>. If the check continues
to fail, exchange the pneumotach and repeat this procedure. If it still fails, please
contact ERT Customer Care.
If any of the strokes are outside this limit, the calibration check will
automatically stop, and the message “Measured volume is outside +/- 3% of
the pump volume.“ will be displayed.
To re-calibrate at any time, click <Calibration Check> on the icon bar on the left
and select the respective icon. Follow the procedures described above.
The calibration check log report can be printed by selecting <Calibration> and the
<Calibration Check Log> icon.