Use Instructions

Page 79/233
MasterScope - Instructions for Use
Version 03.00 • Date 19FEB2020
Proceed as follows:
1. Press the <Calibration Check> icon. In the following screen you will be asked
to verify the current ambient room conditions.
Confirm with the button <Continue>. Following window appears:
2. Use the 3-L calibration syringe for calibration. Connect the calibration syringe
to the wireless spirometry sensor unit.
3. Press <F1>. An automatic zero adjustment of the connected pneumotach will
be performed. The following window appears: