Use Instructions
Page 228/233
MasterScope - Instructions for Use
Version 03.00 • Date 19FEB2020
Calibration Syringe: Volume:
Accuracy:
3 L
+/- 12 mL
Technical Data Calibration Syringe
Technical Data Wireless Spirometer Sensor Unit & Charging
Station
Measuring Range: PEF:
FEV1 and FVC:
0.1 to 16 L/s
0.1 to 8 L
Resolution: PEF:
FEV1 and FVC:
< 5 mL/s
1 mL
Accuracy: PEF:
FEV1 and FVC:
0.1 to 16 L/s: ± 10% of reading or ± 0.3 L/s
0.1 to 8 L: ± 3% of reading or ± 0.050 L
Resistance Spirometer: max. 135 Pa/L/s at 14 L/s
Instantaneous Flow: 0.1 – 14 L/s: +/- 5% or 0.2 L/s
Applied Part: ERT PT, WSSU
Dimensions: Wireless Spirometry Sensor Unit
Charging Station
11 x 8 x 4 cm (L x W x H)
19 x 13 x 7 cm (L x W x H)
Weight: Wireless Spirometry Sensor Unit
Charging Station
0.15 kg
0.27 kg
Medical Power Supply: Model:
Input:
Output:
Cable length:
GTM41134-0606-1.0
100 - 240V 50-60 Hz 0.6 A
5V 1.2A, 6W
1.5 m
Battery: Wireless Spirometry Sensor Unit Built in rechargeable Lithium Ion battery 3.7 V, 640
mAh. Battery will last under standard operating
conditions for about 3 days in standby mode and
2.5 h for constant measurements.
Full charging: 2h
Cycle life: 70% of rated capacity after
500 charging cycles
Protection Class: Charging Station
Wireless Spirometry Sensor Unit
n/a
Internally powered
Mode of Operation: Continuous
Moisture Protection: Power Supply Charging Station
Wireless Spirometry Sensor Unit
IP21
IP20
Measuring Principle: High Quality Pneumotach
Operating Ambient: Temperature:
Relative Humidity:
Barometric Pressure:
+10° C to +35° C
15% to 90%
700 to 1070 hPa
Transport/Storage: Temperature:
Relative Humidity:
Barometric Pressure:
- 20° C to +50° C
15% to 90%
600 to 1200 hPa