Use Instructions

Page 130/233
MasterScope - Instructions for Use
Version 03.00 • Date 19FEB2020
10.2 Performing an ECG Recording
A screen displaying the correct placement of the electrodes for the ECG recording
will appear.
10.2.1 Electrode Test
Before an ECG recording is started, an automatic electrode contact measurement
is performed. This measurement is started by pressing <F1>.
If an electrode contact is indicated as poor, please check the respective contact.
The electrode test program automatically checks whether the electrodes are in
good contact with the skin. If a contact is poor, the respective electrode will
immediately be indicated on the screen and directly at the electrode.
good contact (only on the screen)
bad contact (flashing)
The lead placement check will start automatically as soon as all electrodes are in
good contact with the skin.