Use Instructions

Page 116/233
MasterScope - Instructions for Use
Version 03.00 • Date 19FEB2020
Forced Spirometry: Repeatability of test set
No repeatability: Less than 3 acceptable measurements recorded
• FEV1
repeatability is unacceptable
Difference in FEV1 from best to second best is greater than 150 mL.
FVC repeatability is unacceptable
Difference in FVC from best to second best is greater than 150 mL.
PEF repeatability is unacceptable
Difference in PEF from best to second best is greater than 0.67 L/sec.
Slow Spirometry: Repeatability of test set
IC repeatability is unacceptable
Coefficient of Variation (CV) is greater than 6%.
VC repeatability is unacceptable
Difference in VC from highest to second highest is greater than 150 mL.
Messages for repeatability and/or acceptability are evaluated. This is also
recalculated after every new measurement performed.
E code Description
L IC repeatability is unacceptable
M VC repeatability is unacceptable
N No repeatability: Less than 3 acceptable slow measurements
O Expiration time was too short
P No plateau was detected at the end of the expiration
Q End-expiratory level of the tidal breathing was too variable
Slow Spirometry:
In the MasterScope, acceptability and repeatability errors are implemented as error
codes. These error codes are shown on the screen and on the PFT reports:
Forced Spirometry:
E code Description
A No repeatability: Less than 3 acceptable forced measurements
B FEV1 repeatability is unacceptable
C FVC repeatability is unacceptable
D Expiration time was too short
E No plateau was detected at the end of the expiration
F Back extrapolated volume was too large
G PEF repeatability is unacceptable
H Late peak flow detected
I Coughing was detected in the first part of the expiration