Use Instructions
Page 113/233
MasterScope - Instructions for Use
Version 03.00 • Date 19FEB2020
Screen display after pressing <F7> “Result”.
Explanation of icons from left to right:
• Show expiration only
• Show inspiration only
• Show F/V loop
• Switch to Tiffeneau
• Show quality errors
The quality of the flow-volume curve depends on the subject‘s cooperation.In
order to assess reproducibility (repeatability) and thus the quality of cooperation,
it is recommended to perform at least 3 trials. The result of the best and the
second best trial for FEV1 and FVC may differ by ≤ 150 mL or < 5%.
For FVC ≤ 1 L a difference of ≤ 100 mL is valid*.
* Literature:
- MR Miller et al. Series “ATS/ERS Task Force: Standardisation of Lung Function Testing“, Standardisation
of Spirometry, Eur Respir J 2005; 319-338 Copyright © ERS Journals Ltd. 2005