Owner manual

5www.mastercool.com
6. Use CLEAR, ⇑, ⇓ and shift to input full tank weight.
7. Press ENTER to save the full tank weight. The display will read “GOOD” to confirm.
8. Press ON/OFF button on the monitor and platform to turn unit off.
CALIBRATION PROCEDURE
1. Turn on platform.
2. Turn on monitor. Press and together. Display will then show “SEt-1.”
3. Press TARE/ENTER. Display will show “SEt-11” for kg calibration (for lb calibration press and
“SEt-12” will display.) Press TARE/ENTER to confirm.
4. A “-----” 5 digit number will be displayed, press ⇓.
5. Place calibration weight on scale.
6. Press ⇑. Press TARE/ENTER. Number displayed must match calibration weight. If a digit needs
to be changed, press UNITS/SHIFT to activate digit (digit will blink.) Press the orto change
the number. Repeat until all digits are changed.
7. When the number displayed agrees with the weight on the scale press TARE/ENTER.
Calibration is complete. Monitor will restart automatically.
SPECIFICATIONS:
• Capacity: 243 lbs (110 kg)
Accuracy: [±0.01 lb / ±0.005 kg] or 0.1% of reading whichever is greater.
• Resolution: 0.02 lb / 0.01 kg
Automatic Shut-Off: 20 minutes
• Operating Temperature: 32˚F to 120˚F (0˚C to 49˚C)
• Battery Life: Monitor - approximately 20 hours; platform - approximately 30 hours
REPLACING THE BATTERY:
Replace the 9V battery in monitor or (6) AA batteries in platform when low-battery symbols
prompt on LCD. DO NOT OPERATE WITH A LOW BATTERY.