User Manual

3
www.mastercool.com
TECHNICAL DATA:
Connection:
• Model # 53001: CGA580 (US)
• Model # 53002: DIN477#6 & DIN477#10 (includes adapter to convert
DIN477#6 to DIN477#10 (France/Germany)
• Model # 53003: BS3 (England)
• Model # 53004: AS2473.2 (Australia)
Pressure Regulator:
• 100 - 4500 PSI (7 - 300 Bar) High side
• 30 – 1000 PSI (2 – 70 Bar) Adjustment range
• Integrated Shutoff Valve
Safety Relief Valve: 725 PSI (50 Bar)
Gauge Diameter: 2 5/8 (68 mm) Not including guard
WARNINGS
• Only use with nitrogen gas (N2).
The regulator is not to be used with liquid nitrogen.
• Do not modify the regulator. Doing so could result in personal injury.
• Do not over-pressurize the regulator. This could cause leakage, part damage or
personal injury due to bursting of pressure-containing parts.
• Keep the regulator clean and free of oil.
• Do not use a damaged regulator. Except for replacing the gauges, do not
attempt to repair the regulator.
• Failure to follow instructions can result in personal injury and/or damage to the
equipment.
• Do not install this regulator where service conditions can exceed the
specifi cations of any applicable local, state, or federal codes and regulations.
Wear safety glasses and gloves.
REGULATOR CONNECTION AND USE
Note:
The nitrogen regulator can be used with a charging hose and leak testing
gauge assembly or a charging and testing manifold.
• Make sure the cylinder connection (1) is clean and free from damage.
Turn the T handle (6) all of the way out (counter clockwise, looking from the
bottom).
• Open the output control valve (4) to release any pressure, and then close it.
• Install the nylon gasket or O-ring, if the connection requires one.
• Install the regulator onto the cylinder. Do not over tighten the nut on gasketed
connections.
• Connect the yellow charging hose (11) from the regulator to the leak testing
gauge assembly, making sure the ball valve connection (9) is closed.
(If using a manifold, connect the yellow charging hose (15) of the manifold to
the regulator.)
• Connect the other end of the leak testing gauge assembly (8) to the system.
(If using a manifold, connect the low side hose (14) from the manifold to the
system.)
• Slowly open the cylinder valve. The high side gauge (2) should read the
cylinder pressure.
Adjust the regulator pressure by turning the T handle (6) clockwise until the
required pressure (based on the system’s refrigerant) is shown on the low side
(output) (3) pressure gauge.
NOTE:
This is a non-reliving regulator. To adjust the output pressure lower, some
nitrogen will have to be released from the output (e.g. with no hoses connected,
turn the T handle counter-clockwise, then open the output control valve and
close it. The output gauge should read a lower pressure.)
• If using the leak testing gauge assembly, turn the red needle to the required
refrigerant pressure.