Instruction manual

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 221
mmmmm
Parameters
P927 Distance Measurement
The distance between the surface and the transducer face (displays only as % of Empty).
Use P923 unless the distance information is required in percent.
Master Reset (P999)
This feature resets all parameters to original values.
Use this feature prior to initial programming if arbitrary Parameter Values were used
during a bench test, or after upgrading the software. Following a Master Reset, complete
reprogramming is required.
To perform a Master Reset, access P999 and press CLEAR . C.ALL displays until
the reset is complete.
In dual point units, both points can be reset at the same time by setting the index to 00
and pressing the CLEAR keys.
CAUTION: be careful when using this feature. All data for all points will be reset. For
convenience, be sure to record the values you want to re-enter.
Primary Index Transducer
Values Range: 0.000 to 9999% (Displays as % of Empty) (view only)
Related
P005 Units
P006 Empty
Primary Index
Single Point Model Dual Point Model
Global Transducer
Values See below.