7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 185
mmmmm
Parameters
P703 Emptying Indicator
The empty rate required to activate the LCD Emptying indicator ( ).
This value (in Units (P005) or % of Span (P007) per minute) is automatically set to 1/10 of
the Max Empty Rate (P701).
P704 Rate Filter
Damps Rate Value (P707) fluctuations.
Enter the time or distance interval over which the Rate Value is to be calculated before
the display updates.
This is automatically altered along with Maximum Process Speed (P003).
This value automatically alters the Rate Update Time (P705) and / or Rate Update
Distance (P706). Alternatively, these parameter values may be altered independently.
Primary Index
Single Point Model Dual Point Model
Global Transducer
Values Range: 0.000 to 99.00 m (or equivalent depending on units)
Altered by P003 Maximum Process Speed
Related
P005 Units
P007 Span
P701 Max Empty Rate
Primary Index
Single Point Model Dual Point Model
Global Transducer
Values
0 Rate display not required
Filtered Output
1 Continuously filtered and updated
Interval Output
2 1 minute or 50 mm (2 in)
3 5 minutes or 100 mm (3.9 in)
4 10 minutes or 300 mm (11.8 in)
5 10 minutes or 1000 mm (39.4 in)
Alters P707 Rate Value
Altered by P003 Maximum Process Speed
Related P705 Rate Update Time/ P706 Rate Update Distance