7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 183
mmmmm
Parameters
P664 Temperature
View the transducer temperature in °C.
Value is displayed when is pressed in RUN mode (see
Readings in Run Mode
on
page 23).
If Temp Source (P660) is set to any value other than Fixed Temp, the value displayed is the
temperature measured. If Temp Source is set to Fixed Temp, the P661 value is displayed.
Rate (P700 to P708)
These parameters determine how material level changes are reported.
P700 Max Fill Rate
Adjusts the MultiRanger response to increases in the actual material level (or advance to
a higher Failsafe Material Level, P071).
Enter a value slightly greater than the maximum vessel filling rate. This value, in Units
(P005) or % of Span (P007) per minute, is automatically altered when Maximum Process
Speed (P003) is altered.
Primary Index Transducer
Values Range: -50 to 150 (view only)
Altered By P660 Temp Source
Related
P651 Sound Velocity Calibration
P653 Velocity
P654 Velocity at 20°C
P661 Temp Fixed
Primary Index
Single Point Model Dual Point Model
Global Transducer
Values Range: 0.000 to 99.00 m (or equivalent depending on units)
Altered by P003 Maximum Process Speed
Related
•P005 Units
•P007 Span
P071 Failsafe Material Level
P003 Value Meters/Minute
10.1
2 1
310