Parameters
Flow Records (P320 and P321) [MR 200]
Please note that these parameters apply to the MultiRanger 200 only.
These features are enabled if Operation is set for OCM (P001 = 6), or an OCM device is
defined (P600 0). Use them to identify the occurrence of the record high and low flow
rates as displayed in OCM Max Flow (P604) units, or as a percent of OCM Max Flow.
Press CLEAR keys to reset values once the installation is working correctly.
P320 Flow Max [MR 200]
View the highest flow rate calculated (in units or %).
P321 Flow Min [MR 200]
View the lowest flow rate calculated (in units or %).
LCD Totalizer (P322 and P323) [MR 200]
Please note that these parameters apply to the MultiRanger 200 only.
Use these features to view, reset, or preset the eight-digit display totalizer when
Operation is set for OCM or Pumped Volume (P001 = 6 or 7). The eight-digit totalizer is
divided into two groups of four digits. The four least significant totalizer digits are stored
in P322, and the four most significant digits are stored in P323. Adjust these values
separately to set a new total.
Example
P323 = 0017
P322 = 6.294
Totalizer Display = 00176.294
Primary Index
Single Point Model Dual Point Model
Global Transducer
Values Range: -999 to 9999 (view only)
Related P604 Maximum Flow
Primary Index
Single Point Model Dual Point Model
Global Transducer
Values Range: -999 to 9999 (view only)
Related P604 Maximum Flow