Instruction manual
7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 97
mmmmm
Modbus Register Types
Modbus Register Map
Features affecting
volume
,
mA input
, and
Average or Difference
readings apply to the
MultiRanger 200 only. They are clearly marked.
The memory map of the MultiRanger occupies the Modbus holding registers (R40,001 and
up). This map is used when the protocol is Modbus RTU slave or Modbus ASCII slave.
Register Map for Most Common Data
Legend
Type The type of data held in the group of registers.
Start The first register to hold the referenced data.
Data Type The possible values of the data in the register. See
Data Types
on page
107 for more information.
Description The type of data held in the individual registers.
#R The number of registers used for the referenced data.
Read/Write Indicates whether the register is readable, writeable or both.
Type Description Start
#R
1
1.
Maximum registers shown; fewer may be used depending on options installed.
Data Type
Read/
Write
Word Order 40,062 0/1 R/W
Map ID
Register Map Type 40,063 1 0/1 = P782
R/W
ID
Siemens Milltronics
Product Code
40,064 1 4 = Model 200
6 = Model 100
R
Single Parameter Access (SPA)
R40,090 7 see
Appendix A
on page 114
Point
Data
Reading (3)
2
2.
Varies according to model.
MR 100: Available as reading 1 and reading 2 when in either Single or Dual Point
Mode. In Dual Point Mode, reading 1 and reading 2 are always available.
MR 200: Available as reading 1, reading 2, and Average or Difference when in either
Single or Dual Point Mode. In Single Point Mode, point 2 and 3 are only available if
P001 = Average or Difference. In Dual Point Mode, reading 1 and reading 2 are always
available. Point 3 is only available if P001[3]= Average or Difference.
41,010 2 -20,000 to 20,000
R
Volume (2)
3
[MR 200]
3.
2
nd
volume available in Dual Point Mode only.
41,020 2 -20,000 to 20,000
R