Specifications
tecal portable dri-block
®
calibrators
6
Tecal S and H range Dri-Block Calibrator
This range of temperature calibrators is ideal for calibrating temperature sensors used in the
food and pharmaceutical industries in addition to calibrating any sensors used in the temperature
range outlined in the specifications below.
S Version: Set required Temperature and Calibrate
H Version: Set required Temperature and Calibrate; also programmable for switch test and ramp
rate etc.
Tecal 140S/H
The Tecal 140S provides a complete system for calibrating most types of thermal sensors used
in the temperature range of -40°C to 140°C which includes “K” type thermocouples commonly
used in the food industries, cold storage rooms and environmental monitoring systems.
Where automation in both the field and laboratory are required, the combination of the Tecal
140H with the UPI (see page 13) saves time and money. The user can set the program and leave
the unit to calibrate the probes, automatically storing the results in memory. This allows the user
to attend to other tasks and when convenient download the results to a computer to print the
calibration certificate.
Using advanced peltier technology and state-of-the-art mechanical and electrical components,
these calibrators can rapidly cover the complete temperature range where speed is of the essence
without loss of accuracy or stability. The heater block with a defined measuring zone of 50mm
from the bottom of the block allows for different length/active area sensors. Although the quoted
accuracy and stability is ±0 .2°C & ±0.05°C respectively, one can often achieve an accuracy of
±0.1°C and a stability of ±0.02°C with high class RTDs and thermocouples.
Technical Specification – Note improved Specification 2011
Minimum temperature 45°C below ambient
1
(-60°C cooling)
Maximum temperature 140°C
Temperature accuracy in measuring zone. ±0.2°C
Temperature uniformity in measuring zone ±0.2°C
2
Measuring zone 0 to 50mm from base of well
Temperature stability after 10min ±0.05°C
Display resolution 0.1°C
Heating rate, 20°C to 100°C 5 minutes
Cooling rate, 100°C to 0°C 9 minutes
Programmable ramp rate, °C/min 0.1 to 10, on H version
Switch test on H version
Comms port, 9 way D type Full bi-directional RS-232
Weight 14.0(S), 14.4(H) kg
Dimensions H x W x D, mm 285 x 190 x 426
1
In a maximum ambient temperature of 20°C and when using the CH-5 Chiller a temperature of -40°C can be
achieved.
2
At 100°C
Ordering Information
Product Model Voltage Hz Watts Net
Code Weight (kg)
FDB140SD TECAL 140S 230V 50/60 400 13.5
FDB140SP TECAL 140S 120V 50/60 400 13.5
FDB140SY TECAL 140S 100V 50/60 400 13.5
FDB140HD TECAL 140H 230V 50/60 400 14.0
FDB140HP TECAL 140H 120V 50/60 400 14.0
FDB140HY TECAL 140H 100V 50/60 400 14.0
Accessories
FCALCASE Hard Carry case for Tecal H and S models 8.6kg
140H
See page 9 for full range of inserts available.