Specifications
liquid temperature calibration bathsthermoregulators
Techne
®
invented the ”Clip On” thermoregulator in 1948 and now o ers four “Clip On” units.
Thermoregulators are designed to be used with the Techne
®
unheated water baths or any other
suitable laboratory vessels. They will heat, circulate and safely control the temperature of the liquid
in the bath within precise limits.
TE-10A Tempette
- Temperature range of -20°C* to 95°C
- Excellent temperature stability: ±0.01°C at 40°C
- Simple to use analog control
- Suitable for most routine laboratory applications
- User adjustable over-temperature cut-out for unbeatable safety
TE-10D Tempette
- Temperature range of -40°C* to 120°C
- Excellent temperature stability: ±0.01°C at 40°C
- 4 digit setting with a bright LED digital temperature display
- Suitable for most routine laboratory applications
- User adjustable over-temperature cut-out
- Low liquid level cut-out as standard
Technical Speci cations
Speci cations to DIN 12876 TE-10A TE-10D
Temperature range* -20°C to +95°C -40°C to +120°C
Temperature selection Analog Digital
Temperature stability using water @ 40°C ±0.01°C ±0.01°C
Method of control Proportional PID
Temperature sensor Thermistor PRT
Adjustable over-temperature cut-out Yes Yes
Low liquid level cut-out Yes Yes
Heating/Pumping
Nominal heater power at 120V (W) 1000 1000
Nominal heater power at 240V (W) 1000 1000
Pump capacity liters/minute 10 10
Pump capacity (mbar) 145 145
Dimensions
Extension below base (cm) 14.5 14.5
Dimensions L x W x H (cm) 23.7 x 12.4 x 26 23.7 x 12.4 x 26
* Refrigeration or cooling coil required for near or below ambient operation.
(See Flow and Dip Coolers and the cooling coil).
Cooling coil
Connects to the mains water supply; the water being circulated through the coil should be at
least 5°C cooler than the set of the bath temperature. (Fits to any TE or TU unit).
Ordering information
Product Description
Code
1016100 TE-10A, analog thermoregulator, -20°C to 95°C,
(supplied with clamp), 120V
1016600 TE-10A, analog thermoregulator, -20°C to 95°C,
(supplied with clamp), 240V
1016200 TE-10D, digital thermoregulator, -40°C to 120°C,
(supplied with clamp), 120V
1016700 TE-10D, digital thermoregulator, -40°C to 120°C,
TE-10D
TE-10A
All specifi cations are affected by, but not limited to, the size of the bath, the volume of the item(s) being submersed in the bath and
the ambient of the test area. Please contact Techne for assistance.
15