Specifications
11
temperature measurement products
The CH-5 Chiller Unit
The CH-5 Chiller Unit is designed for use in conjunction with the TECAL 140H Calibrator to achieve
calibration temperatures down to -40°C. Chilled water pumped through the heatsinks on the Tecal
140 units allow the Tecal to reach -40°C.
In normal ambient temperatures of up to 20°C the minimum temperature obtainable with the
TECAL 140 models is 45°C below ambient; by arti cially cooling the heatsink of the units with water
from the chiller -40°C can be obtained.
The unit is designed to run on distilled or deionized water.
The CH-5 can also be used to pass cooling water through the cooling coil on the Calibration Baths
(LCB) and Cooling Probe (7032395); minimum temperature +10°C.
Technical Data
Temperature
Working temperature range 4°C to 15°C
Cooling capacity 400W
Temperature control Thermostat on/o
Electrical control Manual
Dimensions (external)
Length 43 cm
Width 23.5 cm
Height (including pump) 52.4 cm
Bath size:
Length 30 cm
Width 14 cm
Bath depth 15 cm
Bath capacity 5 liters
Pump capacity 20 liters/minute
Pump head 6 meters with water
Net weight 58 lbs
Ordering Information
Product Model Voltage Hz Watts Shipping
Code Weight (lbs)
2040400 CH-5 117 50/60 320 58
CH-5
tecal portable dri-block
®