Root Chapter 10- Specifications
www.masimo.com 88
Masimo
Guidance and Manufacturer's Declaration-Electromagnetic
Immunity
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The ME Equipment is intended for use in the electromagnetic environment specified
below. The customer or the user of the ME Equipment should assure that it is used
in such an environment.
Immunity Test IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment -
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+6 kV
contact
+8 kV air
+6 kV
contact
+8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
+2 kV for
power
supply lines.
+1 kV for
input/output
lines.
-- Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
+1 kV -
differential
mode
+2 kV -
common
mode
-- Mains power quality should be
that of a typical hospital
environment.