User's Manual
Root Chapter 10- Specifications
www.masimo.com 87
Masimo
Guidance and Manufacturer's Declaration-Electromagnetic
Emissions
Guidance and Manufacturer's Declarations - Electromagnetic Emissions
The ME Equipment is intended for use in the electromagnetic environment specified
below. The customer or the user of the ME Equipment should assure that it is used
in such an environment.
Emission Test Compliance Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
Group 1
ME Equipment uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class A
For hospital environment only.
Not intended for use in a domestic environment.
Harmonic
Emissions
IEC 61000-3-2
Class A
Voltage
Fluctuations /
Flicker Emissions
IEC 61000-3-3
Complies