User's Manual
Root Chapter 3- Operation
www.masimo.com 54
Masimo
A Nurse Call can be triggered based on alarm, low Signal IQ events or both. In
addition, Nurse Call Polarity can be inverted to accommodate local Nurse Call
station requirements.
Option Description Factory
Default
Setting
Configurable
Settings
Nurse Call
Polarity
Controls the mechanism of
action for triggering to occur.
Should be changed to
accommodate institutional Nurse
Call settings.
Normal Normal or Inverted
Nurse Call
Trigger
Controls the source of
monitoring which sets off the
trigger.
Alarms
Alarms, Signal IQ
Alarms, and Signal IQ
Port 0 None
None, IAP,
IntelliBridge, and
ASCII 1
Port 1 IAP
None, IAP,
IntelliBridge, and
ASCII 1
SpecTable Radical
Radical, Radical
Module A, Sedline,
and Sedline Numerics
Only
Note: The Nurse Call feature is disabled when Audio Pause is enabled and Nurse
Call Trigger is set to Alarms. For more information about Audio Pause, see Audio
Pause.