User's Manual
Root Chapter 3- Operation
www.masimo.com 53
Masimo
1. Press the
key.
2. When the numeric screen displays, enter the following numbers: 6 2 7 4
Asterisks (****) will be displayed.
To undo an entry, press Backspace.
3. Press Enter to access the password protected screen.
Note: The password will have to be entered every time this screen is accessed.
Option Description Factory
Default
Setting
Configurable
Settings
All Mute
Enabled
Enable parameter Alarm
Silence menu option. See
Sounds.
Off On or Off
Lock Alarm
Volume
Sets the lowest alarm volume
level.
Off 3, 4, or Off
Save as Adult
Saves current profile parameter
as the Adult Profile.
N/A
Press Save to
load.
Save as
Pediatric
Saves current profile parameter
as the Pediatric Profile.
N/A
Press Save to
load.
Save as Neo
Saves current profile parameter
as the Neonatal Profile.
N/A
Press Save to
load.
Factory
Defaults
Options are restored to factory
values.
N/A Press Restore.
Device Output