User's Manual
Table Of Contents
- About this Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1- Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- rainbow Acoustic Monitoring™ (RAM™)
- In Vivo Adjustment™
- Signal IQ® (SIQ)
- Adaptive Threshold Alarm (ATA)
- FastSat® (FST®)
- Sensitivity Modes
- Chapter 2- System Components
- Chapter 3- Setup
- Chapter 4- Operation
- Chapter 5- Alarms and Messages
- Chapter 6- Troubleshooting
- Chapter 7- Specifications
- Measurement Range
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- EMC Compliance
- Safety Standards Compliance
- Radio Compliance
- Guidance and Manufacturer's Declaration- Electromagnetic Emissions
- Guidance and Manufacturer's Declaration- Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 8 - Service and Maintenance
- Appendix
- Index
www.masimo.com 81 Masimo
Chapter 8 - Service and Maintenance
The following chapter contains information about cleaning, battery operation, performance
verification, service, repair, and warranty.
Cleaning
The Radius-7 is a reusable device. The device is supplied and used non-sterile.
The Radius-7 should be cleaned before and after it has been applied to a patient and/or in
accordance with local and governmental regulations to minimize the risk of
cross-contamination.
The Battery Charging Adapter should also be cleaned periodically or according to local and
governmental regulations to minimize the risk of cross-contamination.
CAUTION: Check the enclosure for possible cracks or opening before cleaning.
CAUTION: Do not allow liquids to enter the interior of the device.
The outer surfaces can be cleaned either with a soft cloth dampened with a mild detergent
and warm water solution or they can be wiped down with the following cleaning solutions:
• Cidex Plus (3.4% glutaraldehyde)
• 10% bleach solution
• 70% isopropyl alcohol solution
Using the recommended cleaning solutions on the display panel will not affect the
performance of the Radius-7.
WARNING: Do not attempt to clean or re-use the Armband on multiple patients.
WARNING: Discontinue and dispose of Armband if it appears to be stained or becomes
excessively moist to minimize risk of skin irritation.