User's Manual
Table Of Contents
- About this Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1- Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- rainbow Acoustic Monitoring™ (RAM™)
- In Vivo Adjustment™
- Signal IQ® (SIQ)
- Adaptive Threshold Alarm (ATA)
- FastSat® (FST®)
- Sensitivity Modes
- Chapter 2- System Components
- Chapter 3- Setup
- Chapter 4- Operation
- Chapter 5- Alarms and Messages
- Chapter 6- Troubleshooting
- Chapter 7- Specifications
- Measurement Range
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- EMC Compliance
- Safety Standards Compliance
- Radio Compliance
- Guidance and Manufacturer's Declaration- Electromagnetic Emissions
- Guidance and Manufacturer's Declaration- Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 8 - Service and Maintenance
- Appendix
- Index
Radius-7 Chapter 7- Specifications
www.masimo.com 70 Masimo
Motion [3]
(SpO
2
from 70% to 100%)
Adults, Pediatrics 3%
Low perfusion [4]
(SpO
2
from 70% to 100%)
Adults, Pediatrics 2%
Pulse Rate (PR)
Range 25 to 240 bpm
No motion Adults, Pediatrics 3 bpm
Motion [5] Adults, Pediatrics 5 bpm
Low Perfusion Adults, Pediatrics 3 bpm
Carboxyhemoglobin Level (SpCO) [1]
Range of 1% to 40% Adults, Pediatrics 3%
Methemoglobin Level (SpMet) [1]
Range 1% to 15% Adults, Pediatrics 1%
Total Hemoglobin SpHb [6]
Range of 8 g/dL to 17 g/dL Adults, Pediatrics 1 g/dL
Respiratory Rate (RRa, RRp) [7]
Range of 4 to 70 bpm Adults, Pediatrics 1 breath per minute
Resolution
Parameter Resolution
%SpO
2
1%
%SpCO 1%
%SpMet 0.1%