Chapter 6- Troubleshooting
Troubleshooting Measurements
Symptom Potential Causes Next Steps
Low signal
quality.
Sensor is damaged or
not functioning.
Improper sensor type
or application.
Excessive motion.
Low perfusion.
Verify Sensor type and size and
re-apply sensor. See Directions for Use
for Sensor.
Check and see if blood flow to the site
is restricted.
Check the placement of the sensor.
Re-apply sensor or move to a different
site.
Replace Sensor.
Minimize or eliminate motion at the
monitoring site.
Set to Maximum Sensitivity. See
Sensitivity Modes on page 30 of the
Operator's Manual.
Difficulty
obtaining a
reading.
Interference from line
frequency induced
noise.
Misaligned sensor
Inappropriate sensor or
sensor size.
Excessive ambient or
strobing light.
Excessive motion.
Verify/set 50/60hz menu setting. See
Operator's Manual for Root.
Verify Sensor type and size and
re-apply sensor. See Directions for Use
for Sensor.
Check and see if blood flow to the site
is restricted.
Check the placement of the sensor.
Re-apply sensor or move to a different
site.
Minimize or eliminate motion at the
monitoring site.
Shield the sensor from excessive or
strobing light.