User's Manual
Table Of Contents
- About this Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1- Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- rainbow Acoustic Monitoring™ (RAM™)
- In Vivo Adjustment™
- Signal IQ® (SIQ)
- Adaptive Threshold Alarm (ATA)
- FastSat® (FST®)
- Sensitivity Modes
- Chapter 2- System Components
- Chapter 3- Setup
- Chapter 4- Operation
- Chapter 5- Alarms and Messages
- Chapter 6- Troubleshooting
- Chapter 7- Specifications
- Measurement Range
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- EMC Compliance
- Safety Standards Compliance
- Radio Compliance
- Guidance and Manufacturer's Declaration- Electromagnetic Emissions
- Guidance and Manufacturer's Declaration- Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 8 - Service and Maintenance
- Appendix
- Index
www.masimo.com 65 Masimo
Chapter 6- Troubleshooting
Troubleshooting Measurements
Symptom Potential Causes Next Steps
Low signal
quality.
Sensor is damaged or
not functioning.
Improper sensor type
or application.
Excessive motion.
Low perfusion.
• Verify Sensor type and size and
re-apply sensor. See Directions for Use
for Sensor.
• Check and see if blood flow to the site
is restricted.
• Check the placement of the sensor.
Re-apply sensor or move to a different
site.
• Replace Sensor.
• Minimize or eliminate motion at the
monitoring site.
• Set to Maximum Sensitivity. See
Sensitivity Modes on page 30 of the
Operator's Manual.
Difficulty
obtaining a
reading.
Interference from line
frequency induced
noise.
Misaligned sensor
Inappropriate sensor or
sensor size.
Excessive ambient or
strobing light.
Excessive motion.
• Verify/set 50/60hz menu setting. See
Operator's Manual for Root.
• Verify Sensor type and size and
re-apply sensor. See Directions for Use
for Sensor.
• Check and see if blood flow to the site
is restricted.
• Check the placement of the sensor.
Re-apply sensor or move to a different
site.
• Minimize or eliminate motion at the
monitoring site.
• Shield the sensor from excessive or
strobing light.