User's Manual
Table Of Contents
- About this Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1- Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- rainbow Acoustic Monitoring™ (RAM™)
- In Vivo Adjustment™
- Signal IQ® (SIQ)
- Adaptive Threshold Alarm (ATA)
- FastSat® (FST®)
- Sensitivity Modes
- Chapter 2- System Components
- Chapter 3- Setup
- Chapter 4- Operation
- Chapter 5- Alarms and Messages
- Chapter 6- Troubleshooting
- Chapter 7- Specifications
- Measurement Range
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- EMC Compliance
- Safety Standards Compliance
- Radio Compliance
- Guidance and Manufacturer's Declaration- Electromagnetic Emissions
- Guidance and Manufacturer's Declaration- Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 8 - Service and Maintenance
- Appendix
- Index
Radius-7 Chapter 5- Alarms and Messages
www.masimo.com 64 Masimo
Alarm Message Description Next Step
(Pulse CO-Ox) Sensor
Initializing
Device is checking the sensor for
proper functioning and
performance.
If values are not displayed
within 30 seconds,
disconnect and reconnect
sensor. If values are still
not displayed, replace
with a new sensor.
• (Pulse CO-Ox)
Sensor Off
Patient, or
• (RAM) Sensor
Off Patient
Sensor off patient. Disconnect and reconnect
sensor. Reattach sensor.
Sensor not connected to patient
properly. Sensor is damaged.
Properly reapply the
sensor on the patient and
reconnect the sensor to
the Instrument Module. If
the sensor is damaged,
replace the sensor.