User's Manual

Table Of Contents
Radius-7 Chapter 5- Alarms and Messages
www.masimo.com 62 Masimo
Alarm Message Description Next Step
(Pulse CO-Ox)
Replace
Adhesive
Sensor, or
(RAM) Replace
Adhesive
Sensor
When a single-patient-use sensor is
used, the adhesive portion of the
sensor is non-functional, or the life
of the adhesive portion of the sensor
has expired.
Replace the adhesive
portion of the sensor.
(Pulse CO-Ox)
Replace Cable,
or
(RAM) Replace
Cable
Defects in the Instrument Module Return the device for
servicing.
(Pulse CO-Ox)
Incompatible
Adhesive
Sensor, or
(RAM)
Incompatible
Adhesive
Sensor
Not a compatible Masimo Adhesive
sensor.
Replace with a compatible
Masimo sensor.
Sensor is attached to a device
without an appropriate parameter
installed.
Use a compatible sensor.
Contact your local Masimo
Representative to learn
more about optional
parameter upgrades.
(Pulse CO-Ox)
No Adhesive
Sensor
Connected, or
(RAM) No
Adhesive
Sensor
Connected
When a single-patient-use sensor is
used, the adhesive portion of the
sensor is not connected.
Ensure the adhesive
portion is firmly
connected to the sensor.
(Pulse CO-Ox)
Interference
Detected, or
(RAM)
Interference
Detected
High intensity light such as
pulsating strobe lights, excessive
ambient light sources such as
surgical lights or direct sunlight, or
other monitor displays.
Place a Masimo Optical
Light Shield over the
sensor.
(Pulse CO-Ox) Low
Perfusion Index
Signal too small. Move sensor to better
perfused site.