User's Manual
Table Of Contents
- About this Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1- Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- rainbow Acoustic Monitoring™ (RAM™)
- In Vivo Adjustment™
- Signal IQ® (SIQ)
- Adaptive Threshold Alarm (ATA)
- FastSat® (FST®)
- Sensitivity Modes
- Chapter 2- System Components
- Chapter 3- Setup
- Chapter 4- Operation
- Chapter 5- Alarms and Messages
- Chapter 6- Troubleshooting
- Chapter 7- Specifications
- Measurement Range
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- EMC Compliance
- Safety Standards Compliance
- Radio Compliance
- Guidance and Manufacturer's Declaration- Electromagnetic Emissions
- Guidance and Manufacturer's Declaration- Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 8 - Service and Maintenance
- Appendix
- Index
Radius-7 Chapter 5- Alarms and Messages
www.masimo.com 62 Masimo
Alarm Message Description Next Step
• (Pulse CO-Ox)
Replace
Adhesive
Sensor, or
• (RAM) Replace
Adhesive
Sensor
When a single-patient-use sensor is
used, the adhesive portion of the
sensor is non-functional, or the life
of the adhesive portion of the sensor
has expired.
Replace the adhesive
portion of the sensor.
• (Pulse CO-Ox)
Replace Cable,
or
• (RAM) Replace
Cable
Defects in the Instrument Module Return the device for
servicing.
• (Pulse CO-Ox)
Incompatible
Adhesive
Sensor, or
• (RAM)
Incompatible
Adhesive
Sensor
Not a compatible Masimo Adhesive
sensor.
Replace with a compatible
Masimo sensor.
Sensor is attached to a device
without an appropriate parameter
installed.
Use a compatible sensor.
Contact your local Masimo
Representative to learn
more about optional
parameter upgrades.
• (Pulse CO-Ox)
No Adhesive
Sensor
Connected, or
• (RAM) No
Adhesive
Sensor
Connected
When a single-patient-use sensor is
used, the adhesive portion of the
sensor is not connected.
Ensure the adhesive
portion is firmly
connected to the sensor.
• (Pulse CO-Ox)
Interference
Detected, or
• (RAM)
Interference
Detected
High intensity light such as
pulsating strobe lights, excessive
ambient light sources such as
surgical lights or direct sunlight, or
other monitor displays.
Place a Masimo Optical
Light Shield over the
sensor.
(Pulse CO-Ox) Low
Perfusion Index
Signal too small. Move sensor to better
perfused site.