User's Manual

Table Of Contents
Radius-7 Chapter 5- Alarms and Messages
www.masimo.com 61 Masimo
Messages
The following section lists common messages, their potential causes, and next steps.
Alarm Message Description Next Step
Low battery Battery charge is low. Charge Battery Module by
docking into Battery
Charging Adapter on Root
and powering Root with
AC line power. Replace
Battery Module if
necessary.
Device disconnected Device has lost Bluetooth
connectivity with Root.
Check if Bluetooth is
enabled on Root. See the
Operator's Manual for
Root.
Check connection
between the Instrument
Module and Battery
Module.
Redock the Battery
Module on Root to
re-establish Bluetooth
connectivity.
Speaker failure message Device requires service. Contact Masimo Tech
Support. See Contacting
Masimo on page 84 of the
Operator’s Manual.
(Pulse CO-Ox)
Replace
Sensor, or
(RAM) Replace
Sensor
Reusable sensor has used all of its
available monitoring time
Sensor is non-functional.
Defective sensor.
Replace sensor.
(Pulse CO-Ox)
Incompatible
Sensor, or
(RAM)
Incompatible
Sensor
Not a compatible Masimo sensor. Replace with a compatible
Masimo sensor.
Sensor is attached to a device
without an appropriate parameter
installed.
Use a compatible sensor.
Contact your local Masimo
Representative to learn
more about optional