User's Manual
Table Of Contents
- About this Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1- Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- rainbow Acoustic Monitoring™ (RAM™)
- In Vivo Adjustment™
- Signal IQ® (SIQ)
- Adaptive Threshold Alarm (ATA)
- FastSat® (FST®)
- Sensitivity Modes
- Chapter 2- System Components
- Chapter 3- Setup
- Chapter 4- Operation
- Chapter 5- Alarms and Messages
- Chapter 6- Troubleshooting
- Chapter 7- Specifications
- Measurement Range
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- EMC Compliance
- Safety Standards Compliance
- Radio Compliance
- Guidance and Manufacturer's Declaration- Electromagnetic Emissions
- Guidance and Manufacturer's Declaration- Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 8 - Service and Maintenance
- Appendix
- Index
Radius-7 Chapter 5- Alarms and Messages
www.masimo.com 60 Masimo
Alarm Management
In order to minimize accidental changes to Radius-7's critical settings, alarm management is
restricted to Root.
When Radius-7 is connected to Root, audible alarms will sound on Root but not Radius-7. In
this case, audible alarms can be temporarily silenced on Root. Visual alarms will display on
both Radius-7 and Root until the alarm condition has been addressed. For alarm
management on Root, see Operator’s Manual for Root.
When Radius-7 is not connected to Root, audible alarms will sound on Radius-7. Audible
alarms can be temporarily silenced by touching and holding the Touchpad for 2 seconds.
Visual alarms will continue to display on Radius-7 until the alarm condition has been
addressed.
The following are the factory default settings and configuration options for Alarms:
Option Description Factory
Default
Setting
Configurable Settings on Root
Alarm
Volume
Sets the alarm volume level.
Highest
volume
Slide towards the left to decrease
volume to silence.
Pulse Tone
Volume
Sets the pulse tone volume
level.
Highest
volume
Slide towards the left to decrease
volume to silence.
Audio
Pause
Duration
Sets the length of time that
the audible alarm remains
silenced, when Audio Pause
is enabled.
2 minutes
1, 2, 3 minutes, Permanent*,
Permanent with Reminder*.
If Permanent is selected, there will
be no audible alarms, but visual
alarms will still display.
If Permanent with Reminder is
selected, a tone will sound every
three (3) minutes as a reminder
that Permanent is active.
*Requires user to have All Mute
Enabled in the Access Control
menu. See Operator's Manual for
Root.
Note: In the event of temporary loss of power to Radius-7, the Root will restore alarm setting
to Radius-7 through the re-established Bluetooth connection. If the Radius-7 is used without
a Bluetooth connection to Root, then the alarm settings will be restored to the factory
default.