User's Manual

Table Of Contents
www.masimo.com 59 Masimo
Chapter 5- Alarms and Messages
About Alarms
The Radius-7 visually and audibly indicates alarm conditions that the system detects. Audible
alarms may be silenced, without affecting the operation of visual alarms. See Safety
Information, Warnings and Cautions on page 11.
Alarm Priorities
There are two priorities for alarms:
High
Medium
The following are the audible and visual characteristics for different alarm priorities:
Alarm Priority Audible Characteristics Visual Characteristics
High 571 Hz tone, 10 pulse bursts Flashing Red
Medium 550 Hz tone, 3 pulse bursts Flashing Yellow
The conditions and priorities are provided below:
Parameter
High Limit Alarm Low Limit Alarm
SpO2 Medium High
PR High High
PI Medium Medium
SpCO High Medium
SpMet High Medium
SpHb High High
SpOC Medium High
PVI Medium Medium
RRa High High
RRp High High