User's Manual

Table Of Contents
Radius-7 Chapter 4- Operation
www.masimo.com 56 Masimo
RRp Settings
When using a pulse oximetry or pulse CO-Oximetry sensor with Radius-7, respiration rate can
be determined by the plethysmographic waveform (RRp). This method measures a patient's
respiratory rate based on plethysmographic amplitude changes that correspond to the
respiratory cycle. When using a pulse oximetry or pulse CO-Oximetry sensor, RRp alarms and
RRp settings are active and the parameter label displays as RRp.
Note that the Radius-7 can monitor RRa or RRp but not both simultaneously. RRp is active
under the following conditions:
RRp is installed on the Radius-7.
Pulse oximetry or pulse CO-Oximetry sensor is connected.
Acoustic sensor is not connected.
About
An informational read-only screen appears with an explanation about RRp
Alarms
Option Description Factory
Default
Configuration Options
High
Limit
High Limit is the upper threshold
that triggers an alarm.
30 breaths
per minute
6 to 69 breaths per minute in
steps of 1 breath per minute,
or Off
Low
Limit
Low Limit is the lower threshold
that triggers an alarm.
6 breaths
per minute
5 to 68 breaths per minute in
steps of 1 breath per minute,
or Off
Alarm
Delay
When a High or Low alarm condition
occurs, this feature delays the
audible part of an alarm
30 seconds 0, 10, 15, 30 or 60 seconds
Additional Settings
Option Description Factory
Default
Configuration
Options
Averaging
Time
The length of time over which the system
calculates the average of all data points.
Slow No, Fast, Medium,
Slow or Trending