User's Manual

Table Of Contents
Radius-7 Chapter 4- Operation
www.masimo.com 54 Masimo
RRa Settings
When using an acoustic sensor, respiration rate (RR) is determined by the acoustic (RRa)
signal. See rainbow Acoustic Monitoring™ (RAM™) on page 25 of the Operator's Manual. When
the respiratory rate is determined by the acoustic signal, RRa alarms and RRa settings are
active, and the parameter label displays as RRa.
RRa is active under the following conditions:
RRa is installed on Radius-7.
Acoustic sensor is connected.
About
An informational read-only screen appears with an explanation about RRa
Alarms
Option Description Factory
Default
Configuration Options
High Limit High Limit is the upper threshold
that triggers an alarm.
30 breaths
per minute
6 to 69 breaths per minute
in steps of 1 breath per
minute, or Off
Low Limit Low Limit is the lower threshold
that triggers an alarm.
6 breaths
per minute
5 to 68 breaths per minute
in steps of 1 breath per
minute, or Off
Respiratory
Pause
The duration of time that triggers
an alarm if no breaths are
detected
30 seconds 15, 20, 25,30, 35, or 40
seconds
Alarm Delay When a High or Low alarm
condition occurs, this feature
delays the audible part of an
alarm
30 seconds 0, 10, 15, 30 or 60 seconds
Additional Settings
Option Description Factory
Default
Configuration
Options
Averaging
Time
The length of time over which the system
calculates the average of all data points.
Slow No, Fast, Medium,
Slow or Trending