Radius-7 Chapter 4- Operation
www.masimo.com 52 Masimo
SpMet Settings
About
An informational read-only screen appears with an explanation about SpMet
Alarms
Option Description Factory
Default
Configuration Options
High
Limit
High Limit is the upper threshold that
triggers an alarm.
10
1% to 2%, in steps of 0.1%,
2.5% to 99.5% in steps of
0.5%, or Off
When set to Off, alarm is
disabled
Low
Limit
Low Limit is the lower threshold that
triggers an alarm.
Off 0.1% to 2%, in steps of 0.1%
2.5% to 99%, in steps of
0.1%, or Off
When set to Off, alarm is
disabled.
In Vivo
Option Description Factory
Default
Configuration
Options
Enabled See In Vivo Adjustmenton page 27 of the
Operator's Manual.
Off On or Off
Offset
Amount
See In Vivo Adjustment™ on page 27 of the
Operator's Manual.
0 when
turned On
± 3% in steps of ±
0.1%