User's Manual
Table Of Contents
- About this Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1- Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- rainbow Acoustic Monitoring™ (RAM™)
- In Vivo Adjustment™
- Signal IQ® (SIQ)
- Adaptive Threshold Alarm (ATA)
- FastSat® (FST®)
- Sensitivity Modes
- Chapter 2- System Components
- Chapter 3- Setup
- Chapter 4- Operation
- Chapter 5- Alarms and Messages
- Chapter 6- Troubleshooting
- Chapter 7- Specifications
- Measurement Range
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- EMC Compliance
- Safety Standards Compliance
- Radio Compliance
- Guidance and Manufacturer's Declaration- Electromagnetic Emissions
- Guidance and Manufacturer's Declaration- Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 8 - Service and Maintenance
- Appendix
- Index
Radius-7 Chapter 4- Operation
www.masimo.com 49 Masimo
SpHb Settings
About
An informational read-only screen appears with an explanation about SpHb.
Alarms
Option Description Factory
Default
Configuration Options
High
Limit
High Limit is the upper
threshold that triggers an
alarm.
17.0 g/dL
(11.0
mmol/L)
2.0 g/dL to 24.5 g/dL in steps of 0.1
g/dL, or Off
(2.0 mmol/L to 15.0 mmol/L in steps of
0.1 mmol/L, or Off)
When SpHb Precision is set to 1.0, the
values are rounded to the nearest whole
number.
When set to Off, alarm is disabled.
Low
Limit
Low Limit is the lower
threshold that triggers an
alarm.
7.0 g/dL
(4.0
mmol/L)
1.0 g/dL to 23.5 g/dL in steps of 0.1
g/dL. or Off
(1.0 mmol/L to 14.5 mmol/L, in steps of
0.1 mmol/L, or Off)
When SpHb Precision is set to 1.0,
values are rounded to the nearest whole
number.
When set to Off, alarm is disabled.
In Vivo
Option Description Factory
Default
Configuration
Options
In Vivo
Calibration
See In Vivo Adjustment™ on page 27
of the Operator's Manual.
Off On or Off
In Vivo
Calibration Offset
See In Vivo Adjustment™ on page 27
of the Operator's Manual.
0 when
turned On
± 3 g/dL in steps of ±
0.1 g/dL