User's Manual
Table Of Contents
- About this Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1- Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- rainbow Acoustic Monitoring™ (RAM™)
- In Vivo Adjustment™
- Signal IQ® (SIQ)
- Adaptive Threshold Alarm (ATA)
- FastSat® (FST®)
- Sensitivity Modes
- Chapter 2- System Components
- Chapter 3- Setup
- Chapter 4- Operation
- Chapter 5- Alarms and Messages
- Chapter 6- Troubleshooting
- Chapter 7- Specifications
- Measurement Range
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- EMC Compliance
- Safety Standards Compliance
- Radio Compliance
- Guidance and Manufacturer's Declaration- Electromagnetic Emissions
- Guidance and Manufacturer's Declaration- Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 8 - Service and Maintenance
- Appendix
- Index
Radius-7 Chapter 4- Operation
www.masimo.com 47 Masimo
PI Settings
About
An informational read-only screen appears with an explanation about PI.
Alarms
Option Description Factory
Default
Configuration Options
High
Limit
High Limit is the upper threshold that
triggers an alarm.
Off Step size:
0.04 to 0.09 in steps of
0.01
0.10 to 0.90 in steps of
0.10
1 to 19 in steps of 1, or
Off
Low
Limit
Low Limit is the lower threshold that
triggers an alarm.
Off Step size:
0.03 to 0.09 in steps of
0.01
0.10 to 0.90 in steps of
0.10
1 to 18 in steps of 1, or
Off
Additional Settings
Option Description Factory
Default
Configuration
Options
Averaging
Time
The length of time over which the system
calculates the average of all data points.
Long Short or Long