User's Manual
Table Of Contents
- About this Manual
- Product Description, Features and Indications for Use
- Safety Information, Warnings and Cautions
- Chapter 1- Technology Overview
- Signal Extraction Technology® (SET®)
- rainbow Pulse CO-Oximetry Technology
- rainbow Acoustic Monitoring™ (RAM™)
- In Vivo Adjustment™
- Signal IQ® (SIQ)
- Adaptive Threshold Alarm (ATA)
- FastSat® (FST®)
- Sensitivity Modes
- Chapter 2- System Components
- Chapter 3- Setup
- Chapter 4- Operation
- Chapter 5- Alarms and Messages
- Chapter 6- Troubleshooting
- Chapter 7- Specifications
- Measurement Range
- Accuracy
- Resolution
- Electrical
- Environmental
- Physical Characteristics
- Alarms
- Display Indicators
- EMC Compliance
- Safety Standards Compliance
- Radio Compliance
- Guidance and Manufacturer's Declaration- Electromagnetic Emissions
- Guidance and Manufacturer's Declaration- Electromagnetic Immunity
- Recommended Separation Distances
- Symbols
- Citations
- Chapter 8 - Service and Maintenance
- Appendix
- Index
Radius-7 Chapter 4- Operation
www.masimo.com 46 Masimo
Additional Settings
Option Description Factory
Default
Configuration
Options
Averaging
Time
The length of time over which the system
calculates the average of all data points.
8
seconds
2-4, 4-6, 8, 10,
12, 14 or 16
seconds
FastSat Enable/disable FastSat feature for rapid tracking
of oxygen saturation changes. When enabled, the
averaging algorithm evaluates all saturation
values, providing an averaged saturation value
that is a better representation of the patient's
current oxygenation status.
Off On or Off
PR Settings
About
An informational read-only screen appears with an explanation about PR.
Alarms
Option Description Factory
Default
Configuration Options
High
Limit
High Limit is the upper threshold
that triggers an alarm.
140 bpm 35 bpm to 235 bpm, in
steps of 5 bpm
Low
Limit
Low Limit is the lower threshold that
triggers an alarm.
50 bpm 30 bpm to 230 bpm, steps
of 5 bpm