5.4. Service Messages:
Upon detection of a system error, the main software alerts the
user with an audible alarm indicator and displays the
appropriate error code on the display. The MS board is held in
reset and no other runtime functionality is available to the user
until the unit is power cycled or until power is completely
removed and re-applied to the system (removal and/or
replacement of battery).
System Error System
Error
Code
MS board communication Failure 1
MS board Diagnostics Failure 2
User Interface Waveform Overwrite 3
Sound Processor Waveform Overwrite 4
Main Processor Waveform Overwrite 5
Sat Pulse Rate Update Failure User Interface 6
Sat Pulse Rate Update Failure Sound Processor 7
User Interface One Time Command Overwrite 8
Sound Processor One Time Command Overwrite
9
Main Processor One Time Command Overwrite 10
Vor Voltage Error 11
Vbatt Voltage Error 12
V5 Voltage Error 13
Vcc Voltage Error 14
User Interface Revision Table Mismatch 15
Sound Processor Revision Table Mismatch 16