Service manual

Radical Signal Extraction Pulse Oximeter Service Manual
UM10029–0406
4-5
4.12.5. Repeat steps 4.12.1 to 4.12.3 with the Low and High Pulse
Rate Alarm parameter.
4.12.6. Reset the alarm limits again to the original settings and accept
the changes by pressing the Mode/Enter button.
4.13. Date and Time Setting
4.13.1. Press the Mode / Enter button 4 times and verify that “Set CL”
message appears on the display.
4.13.2. Press the Next arrow key and “Yr XX” will be displayed.
4.13.3. Verify that correct year is displayed.
4.13.4. Use Up / Down arrow buttons to set accordingly.
4.13.5. Press the Next arrow key and “nn XX” will be displayed.
4.13.6. Verify that correct month is displayed.
4.13.7. Use Up / Down arrow buttons to set accordingly.
4.13.8. Press the Next arrow key and “dAY XX” will be displayed.
4.13.9. Verify that correct day is displayed.
4.13.10. Use Up / Down arrow buttons to set accordingly.
4.13.11. Press the Next arrow key and “Hr XX” will be displayed.
4.13.12. Verify that correct hour is displayed.
4.13.13. Use Up / Down arrow buttons to set accordingly.
4.13.14. Press the Next arrow key and “nn XX” will be displayed.
4.13.15. Verify that correct minute is displayed.
4.13.16. Use Up / Down arrow buttons to set accordingly.
4.13.17. Press the Mode / Enter button.
4.13.18. Cycle power and verify the settings.
4.13.19. Turn power off.